21 CFR Part 820 QMSR — FDA-Specific Delta (US Medical Devices)

The residual FDA-specific requirements (MDR/corrections/tracking/UDI cross-references, US-specific records and labeling controls) layered on top of ISO 13485 under FDA's Feb 2026 QMSR rule — use alongside, not instead of, the ISO 13485 rubric.

21 CFR Part 820 QMSR — FDA-Specific Delta Rubric (US Medical Devices)

What this is: On February 2, 2024, FDA published the Quality Management System Regulation (QMSR) final rule (89 FR 7496, Feb. 2, 2024, docket FDA-2021-N-1288), which took effect February 2, 2026 and replaced the old, standalone Quality System Regulation (QSR) text in 21 CFR Part 820. The QMSR works by incorporating ISO 13485:2016 by reference (new §820.10(a), via new §820.7) as the substantive core of US device manufacturing quality requirements, rather than restating quality-system requirements independently the way the old QSR did. Regscribe's iso13485-rubric.md already covers that substantive core in full.

This document is the residual, FDA-only layer that sits on top of ISO 13485 — not a second copy of it. A small number of specific §820 sections were deliberately retained or newly added because FDA judged ISO 13485 alone inadequate or silent on points the US statutory/regulatory scheme requires. A document reviewed only against Regscribe's ISO 13485 rubric would miss every requirement in this file. A document reviewed only against this file, without also being checked against ISO 13485, would miss nearly everything else in Part 820 — the two rubrics are meant to be applied together for a US device manufacturer, never as alternatives.

Why this rubric exists now: Regscribe previously excluded 21 CFR Part 820 entirely from its knowledge base, on the reasoning that QMSR "subsumed" it into ISO 13485. Re-verification this session found that framing overstated the case — the exclusion note itself (and the ISO 13485 rubric's own "FDA QSR → QMSR harmonization" section) already named four FDA-specific retained areas, but never turned them into a scoreable rubric, and the sourcing behind that note was search-snippet-only (NSF/AAMI/BSI pages that 403'd on direct fetch), not primary-source-verified. This document exists to close that gap: primary-source verification of the actual current §820.10/§820.35/§820.45 text, plus enough grounding on the four cross-referenced parts (803/806/821/830) to cite them correctly.


Sourcing note — read before using this in production scoring

What was fetched directly from a primary source this session:

  • eCFR's own Versioner API (ecfr.gov/api/versioner/v1/full/{date}/title-21.xml?part=820) — this is eCFR's structured-data endpoint for the actual current regulatory text of 21 CFR Part 820, fetched for the version dated 2026-07-01 (i.e., the live text as of this session, five months after the Feb 2, 2026 QMSR effective date — not a stale pre-effective-date snapshot). This is a genuine primary-source fetch, distinct from and stronger than the ISO 13485 rubric's FDA QMSR section, which relied entirely on search-result snippets after direct fetches 403'd/404'd. Section text below (§820.10, §820.35, §820.45, and the confirmed presence of new §820.3, §820.7) is drawn from this fetch.
  • A note on quote fidelity: the tool used to fetch and read this content applies a hard per-quote length cap (independent of copyright status — CFR text is public domain), so what appears below is a faithful structural paraphrase, paragraph-by-paragraph and sub-item-by-sub-item, not a verbatim reproduction of the full regulatory text. Every lettered/numbered subparagraph identified in the fetch is represented below; nothing was skipped or invented. Where a short verbatim phrase was returned by the fetch, it's quoted in quotation marks; everything else is paraphrase clearly presented as paraphrase. Before this goes into a customer-facing citation that needs word-for-word regulatory language, re-verify the exact wording directly against eCFR or the printed CFR — this document is reliable on structure, requirement substance, and citation accuracy, but should not be treated as a substitute for the primary text when exact phrasing matters (e.g., a legal dispute).
  • Federal Register final rule (89 FR 7496, Feb. 2, 2024) — direct fetch of federalregister.gov's own page redirected through a bot-protection gateway (unblock.federalregister.gov) that the fetch tool could not get past; govinfo.gov's hosted copy of the same Federal Register document was successfully fetched instead (govinfo.gov/content/pkg/FR-2024-02-02/html/2024-01709.htm) and confirmed: publication date February 2, 2024; effective date February 2, 2026; FR citation Volume 89, Number 23, pages 7496–7525, docket 2024-01709. This is a primary-source confirmation of the rule's existence, dates, and citation — not a snippet.
  • Current-status verification (via WebSearch, multiple independent industry-compliance sources — Greenlight Guru, Morgan Lewis, MedAcuity, Gardner Law, QT9 — cross-checked against each other, not a single source): FDA did not provide any transition/grace period from QSR to QMSR compliance, either in the rulemaking itself or via subsequent enforcement discretion. As of this session (2026-07-30, roughly six months after the Feb 2, 2026 effective date), no source found reports any delay, extension, or further amendment to the rule. The old QSR text has been fully removed from 21 CFR Part 820; FDA's updated device-establishment inspection process (Compliance Program 7382.850) has been in use since the effective date. This is WebSearch-sourced secondary confirmation, not a direct fetch of FDA's own QMSR status page — a direct fetch of fda.gov/medical-devices/postmarket-requirements-devices/quality-management-system-regulation-qmsr was attempted and returned HTTP 404 (the URL may have moved or the page restructured) — flagged as a gap below, same honesty standard as the ISO 13485 rubric's own FDA-page 404s.
  • 21 CFR Parts 803 (MDR), 806 (Corrections and Removals), 821 (Device Tracking), 830 (UDI) — not independently fetched from eCFR this session (out of scope per the task: "just enough to correctly cite them as the §820.10 cross-references require," not full separate rubrics). What appears below for each is drawn from WebSearch results only, cross-checked across 2-4 independent sources per part (a mix of law-firm/consultancy summaries and, for Part 806, a direct quote-bearing summary that appears to draw on the actual regulatory text of §806.10). This is weaker sourcing than the §820.10/820.35/820.45 sections above and is flagged explicitly in each row below — treat the core-requirement summaries for these four parts as directionally reliable (right ballpark for timelines/triggers) but re-verify against eCFR directly before citing a specific paragraph letter/number from Parts 803, 806, 821, or 830 to a customer, the same discipline applied to every other search-snippet-sourced item in Regscribe's rubrics.

Nothing below is fabricated. Every requirement, citation, and paraphrase is traceable to one of the sources above; where sourcing is weak, that is stated in the row itself, not smoothed over.


0. READ THIS FIRST — this is a DELTA rubric, not a full Part 820 rubric

Do not use this document as a standalone Part 820 review. As of February 2, 2026, the substantive core of US device-manufacturer quality-system compliance is ISO 13485:2016 — new §820.10(a) requires manufacturers to "document a quality management system that complies with the applicable requirements of ISO 13485." Regscribe's iso13485-rubric.md is the document that reviews that core. This rubric covers only what's layered on top of that core: the sections FDA explicitly retained, added, or modified because it judged ISO 13485 alone insufficient for the US regulatory scheme. Applying this rubric alone to a device-QMS document and calling it "Part 820-compliant" would produce a dangerously incomplete review — the overwhelming majority of Part 820's actual substance (design controls, CAPA, production controls, purchasing controls, management responsibility, etc.) is inside ISO 13485 now, not restated here.

The four areas covered here, and why each survives ISO 13485 incorporation:

  1. §820.10's cross-references to Parts 803, 806, 821, 830 — ISO 13485 is a generic international standard with no US-statute-specific reporting mechanics; it does not itself require MDR adverse-event reporting, corrections/removals reporting, device tracking, or UDI assignment. These are US-only regulatory obligations that a device manufacturer must meet regardless of ISO 13485 conformance.
  2. §820.35 (Control of records) — layered on top of ISO 13485's own record-control clause (§4.2.5), with US-specific additions: mandatory minimum data elements for complaint records and servicing records, UDI recording requirements, and a confidentiality-marking provision tied to FDA's public-information regulations (21 CFR Part 20).
  3. §820.45 (Device labeling and packaging controls) — FDA explicitly stated it did not find ISO 13485's own labeling/packaging provisions (Clause 7.5.1) adequate, and added supplemental pre-release inspection and mixup-prevention requirements that have no ISO 13485 equivalent.
  4. Statutory/definitional precedence — ISO 13485's Clause 3 definitions (plus ISO 9000:2015's) generally govern under new §820.3, except where 21 CFR 820.3(b) or the FD&C Act §201(h) statutory definitions differ, in which case the US statutory definition controls. This matters for terms like "device" itself, which has FD&C Act-specific legal weight ISO 13485's generic definitions don't carry.

How to use this alongside iso13485-rubric.md: for any US-market device-QMS document, run both rubrics. ISO 13485 covers the substantive quality-system content; this rubric covers the four items above. A document can pass ISO 13485 cleanly and still fail this rubric (e.g., a labeling-control SOP that doesn't document the §820.45(a) pre-release inspection elements), and that failure would be invisible to an ISO 13485-only review.


Rubric legend

Each row: Requirement (plain language / structural paraphrase) → Citation → Evidence that demonstrates compliance → Common failure pattern


Part A — §820.10's cross-referenced requirements (Parts 803, 806, 821, 830)

New §820.10(b) requires manufacturers to comply with additional FDA-specific regulatory requirements beyond ISO 13485 itself, cross-referencing four other parts by number. Sourcing for §820.10 itself is primary (eCFR Versioner API fetch, per the sourcing note above); the core-requirement summaries for the four cross-referenced parts below are WebSearch-sourced and weaker — flagged per row.

A.1 — §820.10(b) itself: the cross-reference clause

Requirement§820.10(b) requires manufacturers to comply with "other applicable requirements" in the CFR beyond ISO 13485, specifically naming four: (1) assign a unique device identifier per Part 830; (2) document traceability procedures per Part 821, where applicable; (3) notify FDA of complaints that meet the reporting criteria of Part 803; (4) handle advisory notices in accordance with Part 806. §820.10(e) states that failure to comply with any applicable requirement renders the device adulterated under federal law, exposing both the device and responsible parties to regulatory action.
Citation21 CFR 820.10(b)(1)–(4), 820.10(e)
Evidence of complianceA QMS document (typically a top-level quality manual or a cross-reference matrix) that explicitly names all four cross-referenced parts and points to the specific internal procedure governing each — not merely a general statement that "applicable FDA regulations are followed."
Common failure patternA QMS that is fully ISO 13485-conformant but treats UDI, tracking, MDR, and corrections/removals as separate compliance programs never cross-referenced from the core quality manual — the four areas exist as real activities somewhere in the organization but the §820.10(b) linkage itself is undocumented, which is itself the gap FDA is checking for.

A.2 — Part 803 (Medical Device Reporting) — core requirement, cited as required by §820.10(b)(3)

RequirementManufacturers and importers must report to FDA any event where information reasonably suggests a device caused or contributed to a death or serious injury, or where a malfunction occurred that would be likely to cause or contribute to a death or serious injury if it recurred. Per WebSearch-sourced summaries (cross-checked across multiple sources), the standard reporting window is 30 calendar days from becoming aware of a reportable event; a 5-work-day expedited report applies where remedial action is needed to prevent an unreasonable risk of substantial harm to public health, or where FDA makes a specific written request. Reports are submitted on Form FDA 3500A via eMDR (mandatory for manufacturers/importers); voluntary reports use Form FDA 3500 (MedWatch).
Citation21 CFR Part 803 (Medical Device Reporting) — not independently fetched from eCFR this session; sourced from WebSearch summaries only, cross-checked across sgsystemsglobal.com, fdaguidelines.com, and complizen.ai. Cited here as the target of §820.10(b)(3)'s cross-reference, not as an independently verified standalone rubric.
Evidence of complianceA documented MDR procedure defining the reportability-assessment criteria (death/serious injury/malfunction-likely-to-recur), the 30-day and 5-work-day timelines, and the eMDR submission mechanism; an MDR event log with assessment-date, submission-date, and reportability-determination fields per event.
Common failure patternNot independently evidenced by primary inspection data this session — flagged as a plausible, industry-summary-supported check rather than one confirmed against real FDA deficiency statistics. A commonly cited industry pattern (unverified against primary FDA data this session) is confusing the 5-work-day expedited trigger (remedial-action-needed or FDA-requested) with a general "serious injury = always 5 days" rule, which the actual criteria don't state.

A.3 — Part 806 (Corrections and Removals) — core requirement, cited as required by §820.10(b)(4)

RequirementPer §806.10(a) (as summarized in the WebSearch fetch, which appears to draw on the actual regulatory text): a manufacturer or importer must submit a written report to FDA of any correction or removal of a device it initiated, if the action was initiated either to reduce a risk to health posed by the device, or to remedy a violation of the FD&C Act caused by the device that may present a risk to health — unless the correction/removal is exempt under §806.1(b) or the information was already provided under §806.10(f). Per §806.10(b), the report must be submitted within 10 working days of initiating the correction or removal. If the same correction/removal is later extended to additional lots/batches, an amended report is required within 10 working days of initiating the extension. The reporting trigger is the manufacturer's own risk rationale for taking the action, not whether an adverse event has actually occurred.
Citation21 CFR 806.10(a), 806.10(b) — not independently fetched from eCFR this session; sourced from WebSearch summaries only (cross-checked across law.cornell.edu/LII, govregs.com, and complizen.ai descriptions, one of which appears to closely track the actual regulatory text of §806.10, but this was not confirmed against a direct eCFR fetch). Cited here as the target of §820.10(b)(4)'s cross-reference.
Evidence of complianceA documented corrections-and-removals procedure that distinguishes reportable corrections/removals (risk-to-health or violation-remedying rationale) from non-reportable ones (routine stock rotation, non-risk-related recalls under the §806.1(b) exemptions), with a 10-working-day report/amendment timeline tracked from the initiation date, not the discovery or completion date.
Common failure patternNot independently evidenced by primary inspection data this session — flagged as a plausible check. A structurally likely failure mode, given the 10-working-day trigger runs from initiation of the correction/removal (not FDA notification or public announcement), is a delayed internal decision-to-report determination eating into the reporting window before the clock is even recognized as running.

A.4 — Part 821 (Device Tracking) — core requirement, cited as required by §820.10(b)(2)

RequirementDevice tracking applies to Class II or Class III devices meeting at least one of three criteria: failure would reasonably likely have serious adverse health consequences; the device is intended to be implanted in the human body for more than one year; or the device is life-sustaining/life-supporting and used outside a device user facility. Per WebSearch-sourced summaries, tracking is not automatic for devices meeting these criteria — since a 1998 change, FDA has discretion to issue a tracking order to a specific manufacturer for a specific device type, rather than tracking applying by default. Where ordered, tracking must extend down to the patient level.
Citation21 CFR Part 821 (Medical Device Tracking Requirements), specifically §821.20 (devices subject to tracking) — not independently fetched from eCFR this session; sourced from WebSearch summaries only, cross-checked across law.cornell.edu/LII and complianceonline.com. Cited here as the target of §820.10(b)(2)'s cross-reference.
Evidence of complianceFor a device plausibly meeting one of the three §821.20 criteria, a documented determination of whether FDA has actually issued a tracking order for that device (not an assumption either way) — and, where a tracking order exists, patient-level tracking records and a documented traceability procedure per §820.10(b)(2)'s own cross-reference wording.
Common failure patternNot independently evidenced this session — flagged as a plausible check. The most likely documentation gap, given tracking is order-triggered rather than automatic, is a QMS that either assumes tracking applies to every device that could plausibly meet the criteria (over-scoping) or never documents the "does a tracking order exist for this device" determination at all (silently under-scoping) — either failure mode is a records gap Regscribe should be able to flag by checking for an explicit tracking-applicability statement.

A.5 — Part 830 (Unique Device Identification) — core requirement, cited as required by §820.10(b)(1)

RequirementEvery medical device label and device package must bear a UDI meeting the requirements of 21 CFR 801 Subpart B and Part 830. A UDI is composed of a mandatory Device Identifier (DI, identifying the specific device version/model and labeler) and, where the label carries lot/batch numbers, serial numbers, manufacturing dates, or expiration dates, a Production Identifier (PI) segment. The UDI must be presented in both plain-text and AIDC (automatic identification and data capture, e.g. barcode) form, issued under an FDA-accredited issuing agency's system conforming to ISO/IEC 15459 series standards. Part 830 also requires submission of device identification data to FDA's Global UDI Database (GUDID).
Citation21 CFR Part 830 (Unique Device Identification), specifically Part 830 Subpart B (requirements for a UDI) and the related labeling cross-reference at 21 CFR 801 Subpart B — not independently fetched from eCFR this session; sourced from WebSearch summaries only, cross-checked across sgsystemsglobal.com and FDA's own guidance-document search snippets. Cited here as the target of §820.10(b)(1)'s cross-reference.
Evidence of complianceA UDI-assignment procedure documenting how DI/PI are generated and issued (naming the FDA-accredited issuing agency used), label/packaging artwork evidencing both plain-text and AIDC UDI presentation, and a GUDID submission record for the device.
Common failure patternNot independently evidenced this session — flagged as a plausible check. A structurally likely gap is a UDI present on the primary device label but not consistently carried through to secondary/shipping packaging, or a PI segment omitted when the label does in fact carry a lot/batch/serial/expiration field that should have triggered one.

Part B — §820.35 (Control of records)

Sourcing: primary (eCFR Versioner API fetch, 2026-07-01 version).

B.1 — Complaint records (§820.35(a))

RequirementManufacturers must maintain records of the review, evaluation, and investigation for any complaints. The fetched text identifies seven required minimum data elements for a complaint record: device name; date the complaint was received; the device's UDI or UPC; complainant name and contact information; the nature and details of the complaint; any correction/corrective action taken; and any reply provided to the complainant. This is layered on top of, not a replacement for, ISO 13485's own complaint-handling clause (§8.2.2).
Citation21 CFR 820.35(a)(1)–(7)
Evidence of complianceA complaint log or complaint-handling record template containing all seven fields as distinct, populated entries — not a free-text narrative that happens to mention some of them. Specifically check for the UDI/UPC field, since this is the field with no ISO 13485 equivalent and the one most likely to be silently dropped from a complaint form built primarily around ISO 13485's own (less prescriptive) requirements.
Common failure patternA complaint-handling SOP built to satisfy ISO 13485 §8.2.2 alone typically captures narrative investigation content well but omits the UDI/UPC field specifically, since ISO 13485 itself doesn't require it — this is the most likely single point where an otherwise-strong ISO 13485 complaint process fails the US-specific §820.35(a) overlay.

B.2 — Servicing records (§820.35(b))

RequirementWhere a manufacturer performs servicing (documented per ISO 13485 Clause 7.5.4), §820.35(b) requires recording, at minimum, six specific elements for each servicing activity: the name of the serviced device; the device's UDI/UPC or other identifier; the date of service; the individual(s) who performed the service; a description of the service performed; and test/inspection data generated during the service.
Citation21 CFR 820.35(b)(1)–(6)
Evidence of complianceService records or a service-call log containing all six specified fields as distinct entries — again, the UDI/UPC identifier field is the one most likely to be missing from a servicing record built only against ISO 13485's own (less specific) Clause 7.5.4 requirement.
Common failure patternServicing records that capture "what was done" narratively but don't independently record the identifying UDI/UPC of the specific serviced unit — a gap that matters because it's exactly the field that would let a complaint or field-safety issue be traced back to a specific serviced device.

B.3 — UDI recording (§820.35(c))

RequirementThe UDI must be recorded for each medical device, or each batch of medical devices, per ISO 13485 Clauses 7.5.1, 7.5.8, and 7.5.9 — this section links the record-keeping requirement specifically to the device identification/traceability clauses already present in the ISO 13485 rubric, but makes the UDI-recording element itself an explicit, separately-cited US requirement.
Citation21 CFR 820.35(c)
Evidence of complianceProduction/batch records (already reviewed under ISO 13485's §7.5.8/7.5.9 in the sibling rubric) additionally confirmed to carry the device's or batch's UDI as a distinct, recorded field — not merely a lot number that happens to also serve as an informal identifier.
Common failure patternNot independently evidenced by inspection data this session — flagged as a structural cross-reference check: a production record that satisfies ISO 13485 §7.5.8/7.5.9's general identification/traceability requirement but never specifically records the formal UDI (as distinct from an internal lot/batch number) technically satisfies the ISO clause while missing this US-specific overlay.

B.4 — Confidentiality marking (§820.35(d))

RequirementRecords a manufacturer deems confidential may be marked to assist FDA in determining disclosability under FDA's public-information regulations (21 CFR Part 20) — a US-specific administrative provision with no ISO 13485 equivalent, since ISO 13485 has no analogue to FDA's own public-records disclosure regime.
Citation21 CFR 820.35(d), cross-referencing 21 CFR Part 20
Evidence of complianceA document-control procedure that includes a confidentiality-marking convention for records the manufacturer wants FDA to treat as protected under Part 20's exemptions, applied consistently to records genuinely containing trade-secret/confidential commercial information (not blanket-marked).
Common failure patternNot independently evidenced this session — a narrow, administrative provision, flagged as a minor completeness check rather than a substantive quality gap; the more consequential failure mode would be never using the marking at all on records that genuinely warrant it, leaving the manufacturer with no documented basis to contest an FDA disclosure decision later.

Part C — §820.45 (Device labeling and packaging controls)

Sourcing: primary (eCFR Versioner API fetch, 2026-07-01 version), cross-checked against a WebSearch-sourced Greenlight Guru summary that independently corroborates the same structure and explicitly states FDA's stated rationale ("FDA did not find ISO 13485's requirements for device labeling and packaging controls to be adequate").

C.1 — Documented labeling/packaging control procedures (chapeau)

RequirementIn addition to ISO 13485 Clause 7.5.1's general production-control requirements, manufacturers must document and maintain procedures describing, in detail, the activities that ensure the integrity, inspection, storage, and operations for labeling and packaging under customary conditions of processing, storage, handling, distribution, and (as appropriate) use of the device.
Citation21 CFR 820.45 (chapeau, preceding (a)–(c))
Evidence of complianceA dedicated labeling-and-packaging-control procedure (distinct from a general production-control SOP) that specifically addresses integrity, inspection, storage, and handling of labeling/packaging materials through the device's real distribution and use conditions.
Common failure patternLabeling/packaging control folded into a general production SOP without the level of specificity FDA's own rationale (labeling/packaging inadequacy under ISO 13485 alone) is aimed at addressing — i.e., a document that references "labeling controls" in passing rather than treating it as its own controlled process.

C.2 — Pre-release examination for accuracy (§820.45(a))

RequirementBefore release or storage, manufacturers must examine labeling and packaging for accuracy against five specific elements: (1) the correct UDI, UPC, or other device identification; (2) the expiration date; (3) storage instructions; (4) handling instructions; (5) any additional processing instructions. This is the section FDA explicitly stated is meant to preserve the substance of the old QSR's §820.120(b) inspection requirement, which FDA judged ISO 13485 did not adequately replace.
Citation21 CFR 820.45(a)(1)–(5)
Evidence of complianceA documented pre-release label/packaging inspection record checking all five elements individually, not a single "labeling verified" sign-off with no element-level detail — this is the single most concrete, checkable requirement in this entire delta rubric, since it maps directly onto a specific, enumerable inspection record.
Common failure patternA generic "QA release" sign-off that doesn't demonstrate which of the five specific elements (UDI/UPC, expiration date, storage instructions, handling instructions, processing instructions) were actually checked — this is exactly the ambiguity FDA's own stated rationale (ISO 13485's labeling requirements judged inadequate) targets, and the most auditable, high-value check in this rubric for Regscribe to flag.

C.3 — Documentation of labeling release (§820.45(b))

RequirementRelease of labeling for use must be documented in accordance with ISO 13485 Clause 4.2.5 (record control) — this section explicitly ties the US-specific pre-release inspection back to the ISO 13485 record-control clause already covered in the sibling rubric, rather than creating an independent record-control standard.
Citation21 CFR 820.45(b), cross-referencing ISO 13485:2016 Clause 4.2.5
Evidence of complianceA labeling-release record satisfying ISO 13485 §4.2.5's general record-control requirements (already scored under the ISO 13485 rubric's Clause 4 section), specifically applied to the labeling-release event itself as a distinct, dated, controlled record.
Common failure patternNot independently evidenced this session — a linking/cross-reference clause rather than a standalone substantive requirement; the practical failure mode is the same one already named for general record control in the ISO 13485 rubric (obsolete/uncontrolled records), just specifically applied to labeling release.

C.4 — Mixup prevention and pre-use inspection (§820.45(c))

RequirementManufacturers must establish procedures to prevent labeling and packaging mixups, including an inspection before use to confirm that labeling and packaging correctly correspond to what's specified in the device's Medical Device File (the ISO 13485 §4.2.3 file already covered in the sibling rubric), with inspection results documented.
Citation21 CFR 820.45(c)
Evidence of complianceA documented pre-use labeling/packaging inspection step (distinct from the pre-release inspection in C.2, which happens earlier in the process) with a recorded outcome, and a mixup-prevention procedure (e.g., physical segregation, barcode verification, or a documented double-check step) that references the specific device's Medical Device File as the source of truth for what the correct labeling/packaging should be.
Common failure patternTreating the pre-release inspection (§820.45(a), at release) and the pre-use inspection (§820.45(c), before use) as the same single check — they are two distinct, separately-required inspection points in the fetched text, and a QMS document that only evidences one of them is missing half of §820.45's actual inspection-point structure.

Part D — Statutory/definitional precedence (§820.3(b) / FD&C Act §201(h))

RequirementNew §820.10(a) generally applies ISO 13485's own Clause 3 definitions (plus ISO 9000:2015's), but §820.3(b) preserves that where those definitions conflict with 21 CFR 820.3(b)'s own definitions or the FD&C Act §201(h) statutory definition (most consequentially, the legal definition of "device" itself), the US statutory/regulatory definition takes precedence.
Citation21 CFR 820.3(b); FD&C Act §201(h) — the precise scope and full list of terms where §820.3(b)/§201(h) actually diverge from ISO 13485 Clause 3 was not independently enumerated this session; this row is confirmed structurally (via the eCFR §820.3 fetch, which confirmed the ISO-definitions-generally-apply-except structure) but the specific set of diverging terms was not itemized — flagged as a follow-up below.
Evidence of complianceNot independently scoreable at this level of detail without the specific divergence list — flag any document whose use of a defined term (e.g., "device," "manufacturer," "component") appears to rely on an ISO 13485-only reading where the US statutory definition might differ, for manual review rather than automated determination.
Common failure patternNot evidenced this session — flagged as a genuine open item rather than guessed at.

Gaps / follow-ups before this goes into production scoring

  1. Parts 803, 806, 821, and 830 were not independently fetched from eCFR this session — as scoped by the task ("just enough to correctly cite them... not full separate rubrics"), the core-requirement summaries in Part A above rely on WebSearch snippets, cross-checked across 2-4 sources each but not primary-source-verified the way §820.10/820.35/820.45 were. If Regscribe's product direction ever needs full standalone rubrics for MDR reporting, corrections/removals, device tracking, or UDI (as opposed to just correctly citing them from within this delta rubric), each would need its own direct-eCFR-fetch pass, following this document's own sourcing bar rather than its Part A bar.
  2. §820.3(b)/FD&C Act §201(h) definitional divergence (Part D) is confirmed structurally but not itemized — the fetch confirmed the "ISO definitions apply except where §820.3(b)/§201(h) differ" structure exists, but did not enumerate which specific terms actually diverge. A follow-up eCFR fetch of §820.3(b) specifically, cross-referenced against ISO 13485:2016 Clause 3's own definitions (which the ISO 13485 rubric itself also doesn't reproduce verbatim, per that document's own copyright-driven sourcing constraint), would be needed to make Part D independently scoreable rather than a manual-review flag.
  3. FDA's own QMSR status/FAQ page returned HTTP 404 to a direct fetch attempt this session (fda.gov/medical-devices/postmarket-requirements-devices/quality-management-system-regulation-qmsr) — current-status confirmation (no delay, no further amendment) rests on WebSearch-sourced industry-compliance summaries (Greenlight Guru, Morgan Lewis, MedAcuity, Gardner Law, QT9), cross-checked against each other but not against FDA's own primary status page directly. Worth a follow-up fetch via a different URL or federalregister.gov's own document-history page for the docket, to confirm no further Federal Register notices have been published against docket FDA-2021-N-1288 since Feb 2024.
  4. No FDA Form 483/Warning Letter data specific to §820.35 or §820.45 deficiencies was sourced this session — unlike some of Regscribe's other FDA-facing rubrics (e.g., the Part 11 rubric's named Warning Letter quotes), the "common failure pattern" rows above are reasoned from the regulatory text and FDA's own stated rationale (the labeling/packaging "not adequate" language), not from real inspection deficiency data. A follow-up pass through FDA's public Form 483/Warning Letter database, filtered to post-Feb-2026 device-establishment inspections citing §820.35/§820.45 specifically, would materially strengthen this rubric's evidentiary grounding once enough post-QMSR inspection history exists to search (the rule is only ~6 months into effect as of this session).
  5. The eCFR fetch used a hard per-quote length cap, so §820.10/§820.35/§820.45 above are faithful structural paraphrase, not verbatim regulatory text — re-verify exact wording directly against eCFR before using any row above in a context where word-for-word phrasing carries legal weight (see Sourcing note above).

Suggested lib/rubrics.ts registry entry (not implemented — flagging for a decision, not doing it unasked)

This rubric is not yet wired into lib/rubrics.ts. If it should become a selectable framework in the product, two changes would be needed:

1. Add a new RubricId union member (e.g. "cfr820-qmsr") and a new RUBRICS entry:

"cfr820-qmsr": {
  label: "21 CFR Part 820 QMSR — FDA-Specific Delta (US Medical Devices)",
  filename: "cfr820-qmsr-delta-rubric.md",
  category: "FDA",
  oneLiner: "The residual FDA-specific requirements (MDR/corrections/tracking/UDI cross-references, US-specific records and labeling controls) layered on top of ISO 13485 under FDA's Feb 2026 QMSR rule — use alongside, not instead of, the ISO 13485 rubric.",
},

2. Remove the existing EXCLUDED_FRAMEWORKS entry for 21 CFR Part 820, whose exact current wording (to remove precisely) is:

{
  label: "21 CFR Part 820 (Quality System Regulation)",
  reason:
    "Effectively subsumed — FDA's Feb 2026 QMSR rule incorporates ISO 13485:2016 by reference into 21 CFR 820, so device manufacturers now need ISO 13485 compliance rather than a separate QSR-only system. See the ISO 13485 rubric.",
  docFilename: "iso13485-rubric.md",
},

This entry's own reasoning is what this session's re-verification found to be an overstatement — it should be removed (not merely edited), since the corrected position is that Part 820 is not fully excluded, it now has a real, scoreable delta rubric of its own.

Both changes touch app code (lib/rubrics.ts) and would need whatever test coverage lib/__tests__/rubrics.test.ts expects for registry changes (per the same precedent noted in the ISO 13485 rubric's own integration note) — flagged, not implemented, since it wasn't asked for in this task.